Independent Suture Tape Internal Brace Reinforcement of Bone-Patellar Tendon-Bone Allografts: Biomechanical Assessment in a Full-ACL Reconstruction Laboratory Model
Authors: Patrick A. Smith, James P. Bradley, John Konicek, Jordan A. Bley, Coen A. Wijdicks
Journal: The Journal of Knee Surgery
Publication year: 2020
DOI: 10.1055/s-0039-1692649
IBISG Summary
This biomechanical study evaluated independent suture tape reinforcement of bone–patellar tendon–bone allografts in a full-construct ACL reconstruction laboratory model. Three groups of ten constructs were tested using different combinations of interference screw fixation, adjustable-loop fixation and adjustable-loop fixation with an independent InternalBrace. Cyclic displacement, construct stiffness and ultimate load to failure were measured.
Key Findings
Adding an independent InternalBrace to the adjustable-loop ACL reconstruction construct reduced cyclic displacement by 31% compared with the same fixation method without suture tape. The reinforced construct also displaced significantly less than the interference screw construct. Adding the InternalBrace did not significantly increase stiffness or ultimate failure load compared with the adjustable-loop construct without augmentation.
Why This Research Matters
Reducing cyclic displacement may help protect an ACL graft from excessive stretching during the vulnerable early healing period. This study provides biomechanical support for independently tensioned suture tape as a load-sharing reinforcement for ACL reconstruction. However, it was a time-zero laboratory study using a porcine-bone and human-allograft model, so it cannot demonstrate improved healing, lower failure rates or better outcomes in patients.