Novel Tape Suture Osteosynthesis for Hyperextension Injuries of the Subaxial Cervical Spine: A Biomechanical Study

Authors: Christopher A. Becker, Titus Kühlein, Sandy Walter, Manuel Kistler, Jan Bruder, Bianka Rubenbauer, Axel Greiner, Wolfgang Böcker, Adrian Cavalcanti Kußmaul

Journal: In Vivo
Publication year: 2023

DOI: 10.21873/invivo.13061

IBISG Summary

This laboratory biomechanical study evaluated a suture tape construct as a motion-preserving treatment for hyperextension injuries of the subaxial cervical spine. AO type B3 injuries were created in 14 synthetic C5–C6 cervical-spine models. Seven were treated using conventional anterior fusion and seven with a tape suture construct. Stability and movement were assessed during extension, flexion, lateral bending and rotation.

Key Findings

The tape suture construct provided extension stability comparable to anterior fusion while allowing significantly greater movement in lateral bending and rotation. No implant failures occurred during biomechanical testing. These findings suggest that the construct may provide sufficient initial stability while preserving more physiological movement than fusion. However, the results were obtained from synthetic laboratory models and do not demonstrate clinical effectiveness in patients.

Why This Research Matters

Anterior cervical fusion stabilises hyperextension injuries but permanently restricts movement at the treated spinal segment and may contribute to complications affecting adjacent levels. A suture tape construct could offer a less rigid, motion-preserving alternative while reinforcing the damaged anterior tension band. This early biomechanical evidence supports further investigation using cadaveric models and clinical studies before the technique can be considered for routine patient treatment.